EGA110 — Definitions & Diagrams
Definitions with Diagrams
Key terms illustrated. Diagrams built from the lecture notes — every label corresponds to an exam target.
Topic 1 — Data & Errors
Accuracy vs Precision — dartboard analogy
Accuracy vs precision — four possible combinations
Gaussian distribution
Gaussian distribution — 68-95-99.7 rule. Shaded area = ±1σ = 68% of measurements.
Absolute error
Eᵢ = |xᵢ − xₜ|. Magnitude of deviation from true value. Same units as measurement.
Relative error
Rᵢ = |xᵢ − xₜ| / xₜ. Dimensionless fraction (or %). Allows comparison across different magnitudes.
Sample standard deviation
s = √[Σ(xᵢ−x̄)²/(N−1)]. Uses N−1 (Bessel's correction). Measures scatter of replicate measurements.
Standard error of the mean
Sₘ = s/√N. Uncertainty in the mean x̄. Decreases with more measurements. NOT the same as s.
Topic 4 — Spectroscopy
Beer-Lambert law — light path through a cuvette
Beer-Lambert law — incident intensity I₀ reduced to I as analyte molecules absorb photons
UV-Vis spectrometer — signal path
UV-Vis spectrometer signal path — source → wavelength selector → sample → detector → output
Molar absorptivity (ε)
Species- and wavelength-specific proportionality constant in Beer-Lambert law. Units: L mol⁻¹ cm⁻¹. Maximum at λ_max.
λ_max
Wavelength of maximum absorbance/molar absorptivity. Most sensitive measurement point; Beer-Lambert most linear here.
Standard addition advantage
Compensates for matrix effects — any species that affects absorbance in the sample also affects the spiked sample equally, so the effect cancels out in the ratio.
PMT (photomultiplier tube)
Detector with cascade of dynodes; each dynode multiplies electrons ~5×; total gain ~10⁶. Range 150–1000 nm. Used in high-sensitivity instruments.
Topic 5 — Titrations
Strong acid / strong base titration curve
Strong acid / strong base titration curve — S-shaped, equivalence point at pH 7, steep jump indicates correct region for indicator selection
EDTA structure — hexadentate ligand
EDTA — hexadentate ligand with 2 N and 4 O donors forming 6 simultaneous coordinate bonds with the central metal ion. Ratio is always 1:1 regardless of metal charge.
Chelate effect
Polydentate ligands form more stable complexes than equivalent monodentate ligands. Driven by entropy — releasing multiple water molecules is thermodynamically favourable.
Volhard method
Back-titration for halides. Add excess AgNO₃ → precipitate AgX → FILTER → add Fe³⁺ indicator → back-titrate excess Ag⁺ with KSCN → red FeSCN²⁺ endpoint.
Filter step is critical — omitting it gives wrong result.
Self-indicating titrant
KMnO₄ is self-indicating. Purple MnO₄⁻ is immediately decolourised by Fe²⁺. First permanent purple = endpoint. No external indicator required.
KMnO₄ : Fe²⁺ stoichiometry
1 MnO₄⁻ + 5 Fe²⁺ + 8 H⁺ → Mn²⁺ + 5 Fe³⁺ + 4 H₂O. Mn goes from +7 → +2 (gains 5e⁻); each Fe²⁺ loses 1e⁻. Ratio 1:5.